Spectrophotometers Ultraviolet Biochemical D-600
Spectrophotometer Model UV/VIS4000
♦ Automatic to measure Absorbance and Transmitance
♦ Compact, space saving
♦ Key board
♦ Can set up various standard curves according to customer solutions and find the concentration of unknowed solutions
♦ Display LCD of HD for results and curves
♦ Memory store of up to 200 results
♦ Coefficient Method to find concentrations
♦ Grating 1200 lines/mm
♦ Automatically WL. Calibration and dark current getting
♦ Can display wavelength, absorbance and transmittance with 5 results per screen
♦ Provided of USB port to connect with a PC through a optional software to display spectrum scanning, kinetics, DNA/Protein calculations and multiwavelength testing results on the screen
Supplied with 4-position 10mm cell changer, externally operable, glass and quartz cuvettes
| Parameters | Features | Parameters | Features |
|---|---|---|---|
| Wavelength range | 190 - 1000nm | Photometric range | -0,300 to 3A;0 to 200%T |
| Measuring modes | ABS/T%/CONC | Detector | Silicon Photodiode |
| Bandwidth pass | 2nm | Display | LCD HD 128*64 |
| Optical system | Single beam. Grating 1200 lines/mm | Light source | Deuterium lamp D2 Halogen lamp |
| Stability | ±0.002 Abs/h (500nm,2h warm up) | Standard cell holder | 4-position 10mm cell changer |
| Wavelength accuracy | ± 1nm | Key board | Membrane |
| Wavelength repeatability | 0,5nm | Interface | USB and Parallel(printer) |
| Wavelength setting | Automatically by keys | Power | AC85-260V 50/60Hz 3A |
| Photometric accuracy | ± 0,5%T | Size | 455x380x180mm |
| Photometric repeatability | 0,3%T | Weight | 11kg |
| Stray light | <0,03%T |
SEE ACCESORIES, CUVETTES AND SPARE PARTS AT END OF CHAPTER OF DOUBLE BEAM SPECTROPHOTOMETERS
| 6.4000.00 | Spectrophotometer UV/VIS. 190-1000nm. Model UV/VIS4000 |
|---|---|
| 6.3000.01 | Optional software for PC.Quantitative/Kinetics/Scanning/Multi WL/Protein/DNA |